Phone : +86 0769-83482055 / +8615338361334 Email: dg@huanyi-group.com
facebook twitter linkedin google+ printerest
You are here: Home > Company News > Two-chamber cold and thermal shock test chamber why oil blockage will occur

Two-chamber cold and thermal shock test chamber why oil blockage will occur

Date: Oct 09,2022 Views: 196

If the operation in the two-chamber cold thermal shock test chamber tilt angle is relatively large, then the test chamber compressor refrigerant oil will flow from the suction tube into the low pressure chamber. When restarting the test chamber equipment compressor refrigeration, the refrigerant oil will be sucked into the cylinder, which can make the compressor load in a short period of time (the compression ratio in the cylinder is based on the volume of gas to determine, however, the liquid is not easy to be compressed), and at the same time pay more attention to is when the refrigerant oil is sent into the condenser, and then through the drying and filtering device into the finer thin tube, it will produce a blockage of the pipe, this The situation is often referred to as the oil plug.





Prev New Next New

Contact us

Telphone: +86 0769-83482055 / +8615338361334

Fax: +86 0769-83482056

Email: dg@huanyi-group.com

Add: Building 2, No. 3, Longkeng Xingye Road, Dongkeng Town, Dongguan,China

google+ twitter facebook printerest linkedin
Copyright 2018 by Dongguan Huanyi Instrument Technology Co., Ltd. All rights reserved. Powered by Onepound
Online Service
Top