Phone : +86 0769-83482055 / +8615338361334 Email: dg@huanyi-group.com
facebook twitter linkedin google+ printerest
You are here: Home > Technical Literature > Method for determination of VOC content in man-made panels in 1m voc environmental climate chamber

Method for determination of VOC content in man-made panels in 1m voc environmental climate chamber

Date: Apr 19,2022 Views: 239

Whether the man-made board meets the requirements of the national standard needs to go through a series of tests, in which the test will use a 1m voc environmental climate chamber. The following is the method for measuring the VOC content of the man-made board for reference.

 

Determination method of VOC release amount of artificial board:

A. Test preparation

Test sample: artificial board

Test equipment: 1m voc environmental climate chamber

Equipment Model: Huanyi Instrument HYV-1000

 

B. Conditions of 1m voc environmental climate chamber:

Temperature: (23±0.5);

Relative humidity: (45±3)%;

Bearing rate: (1.00±0.02)/m;

Air replacement rate: (1.00+0.05)h^-1;

Air velocity on the sample surface: 0.1~0.3 m/s.

 

C. Measurement method:

1. Install the adsorption tube on the thermal desorption instrument, and the airflow direction is opposite to the direction during sampling. The volatile components are desorbed from the adsorbent by heating, and are brought into the cold trap by the carrier gas for pre-concentration: then, the secondary thermal desorption is carried out, and it enters the capillary gas chromatograph through the transfer line. The transfer line temperature is close to the desorption temperature to prevent condensation of the analyte. Select the appropriate desorption time and gas flow rate to ensure that the desorption efficiency of n-hexadecane is better than 95%. Typical analytical conditions are as follows:

Analytical temperature: 260~280℃

Analysis time: 5~15 min

Gas flow rate: 30~50 ml/min

Cold trap maximum temperature: 280℃

Cold trap minimum temperature: ≤-30℃ (depending on the type of cold trap used)

Adsorbent in cold trap: Tcnax TA, 40~100 mg

Transfer line temperature: 220℃

2. Plot the standard curve

3. Sample Analysis 

4. Calculate the results.

The above are the general steps for measuring the VOC content of man-made boards in the 1m voc environmental climate chamber. If you want to know the specific operation method, you can privately message the editor, or visit the official website of "Huanyi Instrument" to consult relevant technical personnel.

                              




Prev New Next New

Contact us

Telphone: +86 0769-83482055 / +8615338361334

Fax: +86 0769-83482056

Email: dg@huanyi-group.com

Add: Building 2, No. 3, Longkeng Xingye Road, Dongkeng Town, Dongguan,China

google+ twitter facebook printerest linkedin
Copyright 2018 by Dongguan Huanyi Instrument Technology Co., Ltd. All rights reserved. Powered by Onepound
Online Service
Top