 Test apparatus 
1 Six stage cascade impactor, the arrangement is specified in Table B.1. 
2 Nebulizer, capable of delivering particles with a mean size of (3,0 ± 0,3) μm when in 
contact with the cascade impactor. 
3 Aerosol chamber, glass, 600 mm long and 80 mm in external diameter. 
4 Flow meters, capable of measuring a flow rate of 28,3 l/min. 
5 Pressure gauge, capable of measuring a pressure of 35 kPa to an accuracy of ± 1 kPa. 
6 Erlenmeyer flasks, 250 ml and 500 ml capacity. 
7 Peristaltic or syringe pump, capable of delivering 0,01 ml/min. 
8 Vacuum pump, capable of maintaining a flow rate of 57 l/min.